Plasty / Hip
Ceramic Bearing in Total Hip Replacement Factors Influencing the Rate of Head Fractures
Quality of Impaction
Performed along the axis of the neck of the prosthesis on a clean Morse taper.
Persistent tissue debris or a damaged Morse taper surface (e.g., scratches) can reduce static fracture load to 20% of its normal value.
Cleanliness of the Morse Taper During Impaction
A clean, undamaged taper is essential to prevent fracture.
Diameter of the Femoral Head
Advantages of Larger Head Diameters:
Reduced risk of impingement between the neck and the rim of the ceramic liner or metal cup.
Increased range of motion.
Decreased risk of dislocation.
Problems with Larger Head Diameters (>36 mm):
Positive offset develops as the head center moves outside the cup center, reducing head coverage angle.
Edge loading occurs during micro-separation phases (inevitable during walking).
Increased friction between bearing couple elements leads to noise (squeaking).
Potential ceramic surface modification, though the impact on fracture rates remains unknown.
Liner Fracture
Contributing Factors:
Implantation Techniques
Cup Design
Functional "Limits" During Edge Loading
Specific Issues with Cup and Liner:
Thin Liners:
Liners that are too thin are more prone to fractures, especially in 50-54 mm diameter cups.
Cup Design:
Cup diameter and thickness indirectly affect liner impaction quality.
Narrow Morse taper angles (<10°) complicate insertion, increasing peripheral stresses.
Recent cups with longer Morse tapers (9 mm or 11 mm) aim to improve liner stability but present challenges with:
Tissue interposition.
Difficulties in cleaning the taper during minimally invasive procedures.
Deformation of perforated, non-impermeable metal cups during impaction.
Insertion Issues:
Misaligned impaction can result in peripheral stress and chipped edges.
Ceramic liners are rigid and cannot adapt to socket deformations, unlike polyethylene liners.
National Joint Registry (NJR) Data: 13-Year Revision Rates in THR
Fixation Type
Bearing Type
Revision Rate (<55 years)
Revision Rate (>55 years)
Cemented
MoP
10
4.5
Cemented
CoP
3.8
4.5
Cemented
CoC
-
-
Uncemented
MoP
7
6
Uncemented
CoP
4.5
4.5
Uncemented
CoC
6.5
5.5
Key Recommendations
Fixation Preferences:
Most patients choose cemented fixation with hard-on-soft bearings.
Bearing Type Preferences:
Ceramic bearings are preferred over metal.
Patients Under 55 Years:
Consider discussing dynamic fixation using uncemented components.
Cup Design Considerations:
Ensure appropriate liner thickness and proper Morse taper alignment to minimize risks of fracture and edge loading.
Related FRCS revision notes Written/reviewed by Kishore Puthezhath
Professor of Orthopaedics and Consultant Paediatric Orthopaedic Surgeon
FRCS (Tr & Orth) revision resource
Reviewed: September 2026
Core revision references: Miller's Review of Orthopaedics; Campbell's Operative Orthopaedics; Orthobullets . Current specialty guidelines are linked within individual notes where applicable.
Prefer studying offline? thinKbox FRCS adds offline personal notes, backlinks, flashcards, spaced review, device-specific capture and private local AI. See the app →